Salter Brecknell WB-521 Series User Manual
Page 68

66
SBI521 Service Manual
Note: When Units = lb:oz, response:
Or
Figure 6.2 Set for 200F
Note: When Units = lb:oz, response:
Or
Command: Z
Response: Zero function is activated (simulate ZERO key). The indicator will not
respond if the scale is in motion, positive overload or negative overload. The indicator
will also not respond if it is not in gross mode or within the zero range specified in the
Setup Menu.
Command: T
Response: TARE function is activated (simulate TARE key). The indicator will not
respond if the scale is in motion, positive overload or negative overload. The indicator
will also not respond if it displaying a negative gross value.
Command: C/U
Response: Changes units of measure (simulate UNIT key),
Command: G
Response: This command is sent to the indicator to revert to gross mode. The indicator
will not respond if the scale is in motion, positive overload or negative overload. The
indicator will also not respond if it is not in net mode.