Bio-Rad Microplate Manager Software User Manual
Page 56
Analyzing Data Using Cutoffs
48
Data transformed – viewed as a Percent (0 - 100 %)
• Low Cutoff: 0
• High Cutoff: 100
Using values from a Standard Curve – standards 0, 5, 10, 25, 50, 100, 250, and 500
defined on template
• Low Cutoff: ! 0
• High Cutoff: ! 500
Using an equation – sample (SAMPLE2) well(s) defined on template
• Low Cutoff: 0
• High Cutoff: (SAMPLE2-0.025)*100
Setting Up a Percent Control Cutoff
Analysis Report
Define the desired cutoffs in the Cutoff Report Editor window.
For example, the Max High is defined as ≥ 85% of the control in all samples. The
other high and low cutoffs are defined as shown:
Identifiers used in
the Cutoff Report
Editor must be used
on the template.