beautypg.com

Appendix 1 pharosfx system specifications, Main component specifications – Bio-Rad PharosFX™ and PharosFX Plus Systems User Manual

Page 39

background image

Appendix 1
PharosFX System Specifications

System Technical Specifications

Specification

Linear dynamic range

1:65535

Pixel resolution

50, 100, 200 or 800 µm, selectable

Image resolution

2 line pairs/mm or 250 µm

Pixel density

16-bit (0–65,535)

Signal decay

32

P 50% retention in 24 hr

Scanning area

35 x 43 cm

Main Component Specifications

Laser Scanner

Specification

Dimensions

69 (D) x 59 (W) x 30.25 (H) cm

Construction

Molded plastic housing, Aluminum mounting plates

Excitation source

532nm Diode Pumped Solid State Laser 15 mW

Weight

30 kg

Electrical
Maximum power

650 Watts

Input voltage range

90–260 VAC, 50–60 Hz

Fuses

No user-serviceable fuses

Operating Requirements

10–32°C, 30–80% humidity

Storage Requirements

0–60°C, 10–90% humidity

External Laser Module Specification

Dimensions

61 (deep) x 31 (wide) x 41 (high) cm

Construction

Welded 16 Gauge steel

Weight

30 kg

Electrical
Input voltage range

100/120 VAC, 50–60 Hz
220/240 VAC, 50–60 Hz

Fuses

6.3 Amp (100/120 VAC)
3.15 Amp (220/240 VAC)

Excitation Source

Option I

488 nM, 10 mW, Ar-ion Laser

Option II

488 nM, 10 mW, Ar-ion Laser and 635 nm,
10 mW Laser Diode

Operating Requirements

10–40°C, 0–90% humidity

Storage Requirements

0–50°C, 0–90% humidity

Screen Eraser-K

Specification

Dimensions

35 (D) x 48 (W) x 57 (H) cm

Construction

Molded plastic

Weight

8.6 kg

Electrical

Input voltage range

100/120 VAC, 50–60 Hz
220/240 VAC, 50–60 Hz

Fuses

6.3 Amp (100/120 VAC)
3.15 Amp (220/240 VAC)

Illumination

4 x 15 Watt user-replaceable fluorescent bulbs

Operating Requirements

10–32°C, 30–80% humidity

Storage Requirements

0–60°C, 10–90% humidity

35