Bio-Rad Experion Protein Analysis Kits User Manual
Page 59
Table C.2. Buffers with mild effects on Experion Pro260 separation or analysis.
Buffer
Concentration
Effect(s)
IEF anode buffer
1.5x: 10.5 mM phosphoric acid
Quantitation
Laemmli buffer
0.5x: 31.3 mM Tris-HCl, pH 6.8, 1% SDS,
Quantitation
12.5% glycerol, 0.005% bromophenol blue,
2.5%
b-mercapotethanol
PSB (ReadyPrep protein extraction kit, 0.5x
Quantitation
membrane I)
RIPA buffer
1x: 150 mM NaCl, 1.0% NP40, 0.5%
sodium deoxycholate, 0.1% SDS, 50 mM Tris
Enlarged system peak shifts to
right; 10 kD sizing affected
TBS
1.5x: 30 mM Tris, 750 mM NaCl pH 7.4
Quantitation
Table C.3. Buffers with severe effects on Experion Pro260 separation or analysis.
Buffer
Effect(s)
Rehydration buffer (ReadyPrep 2-D starter kit)
Enlarged system peak shifts to right; 10 kD peak is missing, sizing affected
S2 (ReadyPrep protein extraction kit, signal)
Enlarged system peak shifts to right; 10 kD peak is missing, sizing affected
Reagent 2 (ReadyPrep sequential extraction kit) Enlarged system peak shifts to right; 10 kD peak is missing, sizing affected
Reagent 3 (ReadyPrep sequential extraction kit) Enlarged system peak shifts to right; 10 kD peak is missing, sizing affected
Total protein extraction buffer
Enlarged system peak shifts to right; 10 kD peak is missing, sizing affected
Criterion XT MES buffer (1.5x)
10 kD peak is missing
Chemicals
Table C.4. Chemicals with no observable effects on Experion Pro260 separation or analysis at or below the
concentrations listed.
Chemical
Concentration
Chemical
Concentration
Acetonitrile
30%
MeOH
15%
Ammonium sulfate
250 mM
Na
3
VO
4
5 mM
Arginine
500 mM
NaCl
250 mM
DMSO
15%
NaF
50 mM
EDTA
10 mM
NaHCO
3
50 mM, pH 8
EGTA
150 mM
NaN
3
0.75%
Ethanol
30%
NaOAc
50 mM, pH 4
Glucose
0.5 M
NaOH
150 mM
Glycerol
35%
NDSB201
1%
Glycine
1 M
Sodium carbonate
150 mM, pH 11.5
Guanidine isothiocyanate
25 mM
Succinic acid
100 mM
HEPES
50 mM, pH 7
Tetramethylammonium chloride
1 M
Imidazole
0.5 M, pH 7
Trehalose
5%
KCl
250 mM
Tris base
1.5 M
MgCl
2
150 mM
Urea
4.5 M
Technical Support: 1-800-4BIORAD • 1-800-424-6723 • www.bio-rad.com
55
Experion Pro260 Analysis Kit