Error
Probable Cause
Recommended Action
Baseline is fluctuating (for
example into very negative
and positive values)
Air bubbles are interfering with electrical
contact in one or more of the wells
Stop the run, remove the chip, and remove and replace
solution in affected wells
Chip is not completely primed
Prime and load a new chip
Samples are too concentrated; protein levels
are out of the range of the assay
Dilute the samples and repeat the analysis
7.3 Contacting Technical Support
If you are having problems with a run and would like the Bio-Rad Technical Support team to review your
data, please submit the following files to LSG_TechServ_US@bio-rad .com for analysis:
n
Run file (C:\Program Files\Bio-Rad Laboratories\Experion Software\Data\ name>\ n Run log (C:\Program Files\Bio-Rad Laboratories\Experion Software\Data\ name>\ n Packet file (C:\Program Files\Bio-Rad Laboratories\Experion Software\Data\ name>\Packet\ n System log (C:\Program Files\Bio-Rad Laboratories\Experion Software\Backup\Version X\ Experion_System.log or .txt) n Lot numbers of the Pro260 chips and the Experion Pro260 analysis kit To export the run file, log file, and packet file, select the run(s) in the tree view and click Export Selected 46 Technical Support: 1-800-4BIORAD • 1-800-424-6723 • www.bio-rad.com Experion Automated Electrophoresis System
Runs. The software exports the files into folders with the corresponding project name(s).