CFX Automation System II

| 33

3

Read this chapter for information about maintaining and troubleshooting the
CFX Automation System II.

n

Maintenance

n

Power Outage

n

Troubleshooting the CFX Automation System II

n

Frequently Asked Questions

Maintenance

Cleaning the CFX Automation System II

Wipe up any spills immediately.

n

Avoid direct contact with spilled liquid

n

Wear protective gloves and safety glasses

The outside surfaces of the plate handler should be cleaned periodically. Clean only the exterior
of the unit. Never remove any instrument covers to clean the inside of the instrument.

n

Use a cloth or sponge dampened with water, alcohol, glass cleaner, or a mild soap solution
diluted with water

n

If using glass cleaner or mild soap, wipe with a cloth or sponge dampened with water after
cleaning to remove any residue

n

Do not use abrasive cleaners

n

Do not spray cleaner onto the instrument

n

Do not allow water or other fluids to drip inside the plate handler

Clean the gripper assembly and the racks with alcohol or other residue-free solvent. The
surface beneath the plate handler racks must be clean at all times. Debris or dust under the
racks will cause the plate handler and the racks to be out of calibration.