Bio-Rad CFX Automation System II User Manual
Page 36

28 |
CFX Automation System II
Chapter 2: Using the CFX Automation System II
28 |
Define How the System Should Respond When Unable to Scan Barcodes or
Match a Barcodes
Table 11 provides an overview of options available for how the system should respond when
unable to scan or match a barcode.
Table 10. Barcode controls and associated system response.
Plate Scanned/Barcode
System Response
Information Required
System Response
Options When Unable
Barcode
Prior to Starting
When Barcode is
to Scan or Match
Controls
Automation Run
Scanned
Barcode (Table 11)
No/No
N/A
N/A
Yes/No
Barcode recorded in the next
Pause and Notify,
plate list row
Skip and Notify,
Run Plate Anyway
Protocol and plate file (or LIMS
or PrimePCR file) in the next
available row will be applied
Yes/Yes
Protocol and plate file (or LIMS or
Pause and Notify,
PrimePCR file) associated with
Skip and Notify
that barcode will be applied to
the plate placed in the CFX System
The displayed order of the plate
list will be modified so that the
currently running plate is moved up
Table 11. System response when unable to scan or match a barcode.
Barcode Controls
Plate Handler Response
Plate List Response*
Email Notifications Sent**
Plate placed in source
Pauses
Yes
rack
Plate placed in waste rack
Continues running with
Yes
the next plate
Plate placed in CFX
Protocol and plate file (or
No
System
LIMS or PrimePCR file)
associated with next row
on the plate list will be
applied to the plate
* If another automation run is executing using a different plate list, that run will continue.
** Email addresses must be properly configured in CFX Manager Software in order for these emails to be received.