3 installing the adsorption cartridges, Installing the adsorption cartridges – Metrohm 850 Professional IC Anion MSM-HC MCS Prep 1 User Manual
Page 76
3.18 Metrohm CO
2
suppressor (MCS)
■■■■■■■■■■■■■■■■■■■■■■
66
■■■■■■■■
850 Professional IC – Anion MSM-HC – MCS – Prep 1
3
Aspirating capillary
For aspirating air low in CO
2
(as a result of
CO
2
4
PEEK pressure screw, long 6.2744.090
5
Capillary connection
6
Pressure screw, short 6.2744.070
Mounted on the air aspirating capillary.
7
Luer coupling 6.2744.120
Mounted on the air aspirating capillary with
pressure screw 6.2744.070.
Connecting the MCS
1 Connecting the MSM
Connect the eluent output capillary (labeled with Detector) using a
6.2744.090 long PEEK pressure screw (34-4) to the MCS inlet (34-1).
2 Connecting the detector
Connect the detector input capillary (36-3) using a long PEEK pres-
sure screw 6.2744.090 (34-4) to the MCS output (34-2).
Caution
If the MCS is not used, the input and output should be sealed with
6.2744.220 stoppers.
3.18.3
Installing the adsorption cartridges
For effective CO
2
removal, the air sucked through the degassing cell
should be as low in CO
2
as possible. To achieve this, the air is aspirated
through a 6.2837.000 CO
2
adsorption cartridge (35-4).
Moisture can block the CO
2
adsorption cartridge. In order to prevent this,
a 6.2837.010 H
2