Page 40

background image

3.6 Rear of the instrument

■■■■■■■■■■■■■■■■■■■■■■

30

■■■■■■■■

850 Professional IC – Anion MSM-HC – MCS – Prep 1

Note

Up to two detectors can be positioned and connected.

Caution

The instrument must be turned off when connecting a detector.

1 Removing rear panel

Unscrew knurled screws (9-1) on the rear panel.

Remove handle if still fastened to the instrument.

Remove rear panel (9-2).

2 Positioning the detector

Put the detector through this opening and position it on the sup-
port surface intended for this purpose. Slide it right up to the
front.

3 Replacing rear panel

Insert the detector cable in a cable feed-through (9-3) on the rear
panel (9-2)
.

Place the detector output capillary in a suitable capillary feed-
through.

Replace (9-2) rear panel.
(Optionally, the handle can be remounted higher up and used as
a holder for MPaks.)

Tighten (9-1) knurled screws.

4 Connecting the detector

Note

The instrument has two detector connection sockets (9-5), Detec-
tor 1
and Detector 2. You must ensure that the selected connec-
tor corresponds to the connector entered in the MagIC Net
method.

Recommendation: Use Detector 1 as standard. In the AnCat sys-
tem with 2 detectors: Detector 1 for anions, Detector 2 for cati-
ons.