1.1 Instrument description

■■■■■■■■■■■■■■■■■■■■■■

2

■■■■■■■■

850 Professional IC – Anion MSM-HC – MCS – Prep 1

used for the purpose of protecting other sensitive components against
contaminations in the solutions used. The fine 2 µm material of the readily
and easily replaceable filter platelets removes particles such as bacteria
and algae from the solutions.

Pulsation damper
The pulsation damper protects the separation column from damage
caused by pressure fluctuations when switching the injection valve, and
reduces interfering pulsations during highly sensitive measurements.

Sample degasser
The sample degasser removes gas bubbles and disolved gases from the
sample. For degassing, the sample flows into a vacuum chamber through
a special fluoropolymer capillary.

Injection valve
The injection valve connects the eluent and sample path through rapid
and precise valve switchover. A precisely measured amount of sample
solution is injected and rinsed with eluent onto the separation column.

Column thermostat
The column thermostat controls the temperature of the column and elu-
ent channel and thus ensures stable measuring conditions. It provides
space for 2 separation columns.

High Capacity Metrohm Suppressor Module (MSM-HC)
The MSM-HC is used for chemical suppression with high capacity during
anion analysis. It is pressure-stable, robust and resistant to solvents.

Peristaltic pump
The Peristaltic pump is used for pumping sample and auxiliary solutions. It
can rotate in both directions.

Metrohm CO

2

suppressor (MCS)

The MCS removes the CO

2

from the eluent flow. This reduces the back-

ground conductivity, improves the detection sensitivity and minimizes the
injection and carbonate peaks.

Conductivity detector
The conductivity detector continuously measures the conductivity of the
liquid passing through and indicates these signals in digital form (DSP –
Digital Signal Processing). The conductivity detector exhibits outstanding
thermal stability and thus guarantees reproducible measuring conditions.

Separation column
The intelligent separation column is the heart of the ion chromatographic
analysis. It separates the different components corresponding to their