7 Technical specifications

850 Professional IC – Anion MSM-HC – MCS – Prep 1

■■■■■■■■

117

Pressure range

Pump

0…50.0 MPa (0…500 bar)

Pump head

0…35.0 MPa (0…350 bar) (applies for the standard PEEK pump head)

Residual pulsa-
tion

< 1 %

Safety shutdown

Function

Automatic shutdown upon reaching the pressure limit values

Maximum pres-
sure limit

Adjustable from 0.1…50 MPa (1…500 bar)

The pump is automatically shut down at the first piston stroke
above the maximum limit value

Minimum pres-
sure limit

Adjustable from 0…49 MPa (0…490 bar)

The shutdown mechanism is inactive at 0 MPa

The shutdown mechanism only becomes active 2 minutes after sys-
tem start

The pump is automatically shut down after 3 piston strokes below
the minimum pressure limit

Gradient capacity

Isocratic or gradient (extendable to quaternary)

Profile

Step, linear, convex and concave

Resolution

< 1 nL/min flow increments

7.8

Sample degasser

Material

fluoropolymer

Resistance to sol-
vents

No restriction (apart from PFC)

Build-up time for
the vacuum

< 60 s

7.9

Injection valve

Actuator time

typ.100 ms

Max. operating
pressure

35 MPa (350 bar)

Material

PEEK