3.8 Eluent

■■■■■■■■■■■■■■■■■■■■■■

40

■■■■■■■■

850 Professional IC – Anion MSM-HC – MCS – Prep 1

Fasten the bottle attachment (16-2) on the eluent bottle (18-10).
The aspiration filter (18-6) must rest on the base of the eluent
bottle.

6 Mounting the adsorber tube

Note

In the case of alkaline eluents and eluents with lower buffer
capacity, the eluent bottle must be equipped with a CO

2

adsorber

(18-4).

First, place a piece of wadding (18-3), then the CO

2

adsorber

(18-4) in the large opening of the adsorber tube (18-2) and close
with the plastic cover.

Fasten the adsorber tube (18-2) using the SGJ clip (18-12) onto
the bottle attachment (18-11).

1

1

2

3

4

5

6

7

8

9

10

11

12

13

14

Figure 18

Eluent bottle – connected

1

Eluent aspiration tubing 6.1834.080
For aspirating the eluent. Pre-installed.

2

Adsorber tube 6.1609.000

3

Wadding

4

CO

2

adsorber

Adsorbs CO

2

from the air (e.g. Merck soda

lime with indicator, no. 6839.10).