Figure 16, Installing tubing weighting and aspiration filter, Figure 17 – Metrohm 850 Professional IC Anion MSM-HC MCS Prep 1 User Manual
Page 49: Eluent aspiration tubing fully equipped
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
850 Professional IC – Anion MSM-HC – MCS – Prep 1
■■■■■■■■
39
4 Installing tubing weighting and aspiration filter
1
2
3
4
5
Figure 16
Installing tubing weighting and aspiration filter
1
Eluent aspiration tubing 6.1834.080
2
Eluent bottle attachment 6.1602.160
3
Tubing weighting
From accessory set 6.2744.210.
4
Clamping screw
From accessory set 6.2744.210.
5
Aspiration filter 6.2821.090
With filter holder from accessory set
6.2744.210.
■
Slide the tubing weighting (16-3) onto the eluent aspiration tub-
ing (16-1).
■
Slide the clamping screw (16-4) onto the eluent aspiration tubing
(16-1).
■
Insert eluent aspiration tubing (16-1) into the aspiration filter
(16-5). The end of the tubing must touch the base of the filter.
■
Screw together clamping screw (16-4) and filter holder (15-1).
The end of the tubing must still touch the base of the filter.
Figure 17
Eluent aspiration tubing fully equipped.
5 Mounting eluent aspiration tubing to the eluent bottle
■
Insert the eluent aspiration tubing into the eluent bottle (18-10).