5 Notes – Maintenance – Faults

817 Bioscan

62

Signal "overload"

• Wrong IC-Net parameter

'

Range

' and '

Fullsca-

le

'

• Damaged reference

electrode

• Damaged working

electrode

• Flow cell is not con-

nected properly

• Wrong measuring

potential

• Check settings


• Check and replace

reference electrode

• Replace working electrode

• Check cable connections

(REF: black, WE: red,
AUX: blue)

• Optimize measuring

potential

Considerable
pressure drop

• Leak in system

• Check connections and

make leakproof

Considerable
pressure rise

• Contamination of the

filter in the 6.2821.100
Filter unit PEEK

• Change of column

packing by injection of
contaminated samples



• Mounting screws

36

of

flow cell overtighten;
25 µm spacer has been
used

• Replace the 6.2821.110

Filter

• Regenerate the column

(see section 5.1.1) or
replace column

Note:
Samples should always
be microfiltered.

• Check screws of flowcell;

use 50 µm spacer

Chromatograms
with poor resolu-
tion, change in the
retention times

• Deterioration in

separation efficiency of
the IC column

• Old Eluent
• Ionic strength or pH of

the sample differs
strongly from eluent

• Regenerate the column

(see section 5.1.1) or
replace column

• Replace eluent
• Dilute sample or change

pH of sample

Extreme peak
broadening,
splitting (double
peaks)

• Dead volume at the

column ends

• Dead volume in IC

system

• Replace column

• Check connections

No signal from
detector

• No mains current

• 817 Bioscan not

switched on

• Faulty fuse

• Incorrect mains voltage

• Measuring cell not

connected

• Dirty working electrode

• Connection to IC Net is

not working properly

• Check mains supply and

voltage (see Section 2.2)

• Switch on the 817 at the

rear panel

• Replace fuse (see section

2.2.1)

• Check mains voltage (see

Section 2.2.1)

• Check connection

• Clean working electrode

(see Section 5.2.3)

• Check connection