3 Operation
817 Bioscan
38
Cell
Manual control of amperometric detector
cell:
Switches on detector cell.
Switches off detector cell.
Attention
Switch on the detector cell only if a conduct-
ing eluent is being pumped through it con-
tinuously!
Oven
Manual control of column oven (heater):
Switches on oven.
Switches off oven.
Temperature
Temperature of column oven.
Range:
15 ... 60 °C
(> room temp. + 10 °C)
Control
Manual setting of 4 remote lines.
Sets remote lines to selected states.
Aux1
,
Aux2
,
Cont1
,
Cont2
Setting of 4 remote lines.
Selection:
0
,
1
,
*
Valve
Manual switching of the injection valve,
e.g. at the 812 Valve Unit:
Switches injection valve to "Fill" position
(control line
Cont1
is set to 0).
Switches injection valve to "Inject" position
(control line
Cont1
is set to 1).
Analog output
Manual setting of the analog output signal
to zero. For RS232 data the full-scale range
(see next page) is shifted until the signal is
located in the center of this range again (s.
also Chapter 6.1).
Triggers the autozero function.
Actual
Display of actual values.
Mode
Display of current detector mode.
Cell
Display of current cell state.
Oven
Display of current oven state.
Control
Display of current control line settings.