5.2.4

Changing separating columns

Identical separation system
If you wish to replace an IC separating column by a column of the same
type, proceed as follows (see Fig. 3):

1 Remove old column

• Switch off flowcell.
• Switch off pump drive of the 709 IC Pump.
• Unscrew connection capillary 21 to flowcell from the column.
• Unscrew preheating capillary 31 from column.

2 Connect new column to injector

• Remove end caps from column.
• Screw preheating capillary 31 to inlet end of separating

column (note flow direction).

3 Rinse column

• Place beaker beneath the column outlet.
• Switch on 709 IC Pump and rinse column with eluent for ca.

10 min with 0.3 mL/min and for ca. 50 min. with 1.0 mL/min,
then switch off pump.

4 Connect column to flowcell

• Screw connection capillary 21 to flowcell to outlet end of

separating column.