2.4 System for sugar analysis – MIC-8
817 Bioscan
21
2.4.2
Settings in «IC Net 2.1»
1 System
As in this configuration only the 817 Bioscan is to be directly
controlled by IC Net then only this instrument is to be installed as
the detector with the help of the system assistant
New system
wizard
.
2 Method
You should first select the standard method supplied for use
with your particular column (e.g. for Metrosep Carb 1:
CARB1_250.mtw
). You can save any alterations under a different
file name.
3 Data source
Select:
Cell( 817 BioScan)
as the signal source.
4 Save system
Then save the system (e.g.
817_MIC-8.smt
).
The following system window should now appear:
5 Connect system and workplace
Under
SYSTEM / Control / Connect to workplace
connect the
new system with the workplace.
At the same time the symbol
will appear in the symbol
bar.
6 Check measuring cell
If the 817 Bioscan is switched on without the flow-through cell
being constantly rinsed by a conducting eluent (e.g. 0.1 M
NaOH) then the measuring cell must not be switched on!
You should now check this in the instrument window of the 817
Bioscan (accessed with a double-click with the mouse on in-
strument symbol) under
Manual / Cell
.
7 Check start mode
Make sure that in this system the 817 Bioscan is standing under
Start with determination
under
Setup / Start mode
.
.