2 status diagram of the remote interface, 3 system initialization, Status diagram of the remote interface – Metrohm 885 Compact Oven SC User Manual
Page 72: Figure 23, Remote status diagram
10.3 System initialization
■■■■■■■■■■■■■■■■■■■■■■
62
■■■■■■■■
885 Compact Oven SC
Outputs
Open Collector
tp
t
p
> 200 ms
active = low, inactive = high
I
C
= 20 mA, V
CEO
= 40 V
+5 V: maximum load = 20 mA
10.2.2
Status diagram of the remote interface
[START]
Output 0
Output 1
Output 2
Output 3
Output 4
Sample ready
Temp ok
Flow on
Error/Stop
Continue
CondOK
Input 0
Input 1
StartSeq
EndSeq
SampleSeq
Figure 23
Remote status diagram
10.3
System initialization
In very rare instances, it could happen that a faulty file system (e.g.
because of a program crash) will lead to an impairment of program func-
tioning. The internal file system needs to be initialized in such cases.
Caution
All user data (methods, solutions, etc.) are deleted if a system initializa-
tion is carried out. Afterwards, the instrument will have the factory set-
tings again.
We recommend creating a backup copy of the system at regular inter-
vals in order to avoid data losses.
After a system initialization the program versions and language files do
not have to be reloaded. Only the selection of the dialog language may
have to be reset in the system settings.