10.3 System initialization

■■■■■■■■■■■■■■■■■■■■■■

62

■■■■■■■■

885 Compact Oven SC

Outputs

Open Collector

tp

t

p

> 200 ms

active = low, inactive = high

I

C

= 20 mA, V

CEO

= 40 V

+5 V: maximum load = 20 mA

10.2.2

Status diagram of the remote interface

[START]

Output 0

Output 1

Output 2

Output 3

Output 4

Sample ready

Temp ok

Flow on

Error/Stop

Continue

CondOK

Input 0

Input 1

StartSeq

EndSeq

SampleSeq

Figure 23

Remote status diagram

10.3

System initialization

In very rare instances, it could happen that a faulty file system (e.g.
because of a program crash) will lead to an impairment of program func-
tioning. The internal file system needs to be initialized in such cases.

Caution

All user data (methods, solutions, etc.) are deleted if a system initializa-
tion is carried out. Afterwards, the instrument will have the factory set-
tings again.

We recommend creating a backup copy of the system at regular inter-
vals in order to avoid data losses.

After a system initialization the program versions and language files do
not have to be reloaded. Only the selection of the dialog language may
have to be reset in the system settings.