7.1 Automation
■■■■■■■■■■■■■■■■■■■■■■
52
■■■■■■■■
885 Compact Oven SC
Only when 'Gas type' = 'other'.
Range
0.001 ... 2.000 (Increment: 0.001)
Default value
1.000
Table 1
Measuring factors of different gases
Gas type
Measuring
factor
Argon
0.950
Helium
0.300
Oxygen
1.000
Air
1.000
Nitrogen
1.000
End of series
Behavior of the instrument at the end of the sample series.
Selection
Stop | Conditioning
Default value
Stop
Stop
After the last sample, the entire system is conditioned one more time.
Afterwards, the needle is moved out of the conditioning vessel. The
gas flow and the oven heating are switched off. The titrator is stopped.
Conditioning
After the last sample, the entire system is conditioned one more time.
Afterwards, the needle is moved out of the conditioning vessel. The
gas flow and the oven heating are switched off. The titrator is kept in
conditioned status.
Conditioning timeout
This setting determines the maximum waiting time for conditioning. If no
stable conditioning signal is obtained from the titrator during this waiting
time for a total duration of 60 s, then an error message will be displayed,
after which the sample series will be stopped.
Range
2 ... 120 min (Increment: 1)
Default value
30 min