beautypg.com

6 instrument diagnosis, 1 loading program versions and language files, Instrument diagnosis – Metrohm 885 Compact Oven SC User Manual

Page 59: Loading program versions and language files

background image

■■■■■■■■■■■■■■■■■■■■■■

6 System settings

885 Compact Oven SC

■■■■■■■■

49

Note

In case of communication problems, try the software handshake (soft-
ware
).

6.6

Instrument diagnosis

6.6.1

Loading program versions and language files

Menu

System Diagnosis

New program versions or language files can be loaded from a USB stick.
The corresponding file must be saved on the USB stick in a directory with
the instrument number (e.g. 848 or 863).

You can distinguish between language files and program files by noting
how the file name is constructed.

Program files
They are instrument-specific. The file name has the following structure:

5XXXyyyy.bin where

XXX =

yyyy =

Instrument type (e.g. 848 for the 848 Titrino plus)

Program version

Language files
They can be recognized by means of the two-digit language code in the
file name. A language file contains the dialog texts for various instrument
types. It is not instrument-specific. The file name has the following struc-
ture:

5848xxxxYY.bin where

xxxx =

YY =

Version number

Language, e.g. DE (German), FR (French), ES (Spanish)

Loading a file

Proceed as follows:

1 Connect the USB stick

Plug in the USB stick with the 6.2151.100 adapter (USB MINI
(OTG) - USB A) at the USB connector on the instrument.

Switch on the instrument.