6 instrument diagnosis, 1 loading program versions and language files, Instrument diagnosis – Metrohm 885 Compact Oven SC User Manual
Page 59: Loading program versions and language files
■■■■■■■■■■■■■■■■■■■■■■
6 System settings
885 Compact Oven SC
■■■■■■■■
49
Note
In case of communication problems, try the software handshake (soft-
ware).
6.6
Instrument diagnosis
6.6.1
Loading program versions and language files
Menu
▶ System ▶ Diagnosis
New program versions or language files can be loaded from a USB stick.
The corresponding file must be saved on the USB stick in a directory with
the instrument number (e.g. 848 or 863).
You can distinguish between language files and program files by noting
how the file name is constructed.
Program files
They are instrument-specific. The file name has the following structure:
5XXXyyyy.bin where
XXX =
yyyy =
Instrument type (e.g. 848 for the 848 Titrino plus)
Program version
Language files
They can be recognized by means of the two-digit language code in the
file name. A language file contains the dialog texts for various instrument
types. It is not instrument-specific. The file name has the following struc-
ture:
5848xxxxYY.bin where
xxxx =
YY =
Version number
Language, e.g. DE (German), FR (French), ES (Spanish)
Loading a file
Proceed as follows:
1 Connect the USB stick
■
Plug in the USB stick with the 6.2151.100 adapter (USB MINI
(OTG) - USB A) at the USB connector on the instrument.
■
Switch on the instrument.