8 Carrying out a determination
885 Compact Oven SC
■■■■■■■■
55
It is advantageous for air bubbles to be stirred into the working
medium and forcefully broken up in the process. The air in the KF
titration cell must also be dried.
Creating a sample table at the titrator
1 Create a sample table on the KF titrator. For each blank, insert a line
which calls up a method for the determination of a blank.
You will find the necessary settings for the method in the KF titrator
manual.
Note
It is imperative that an Extraction time of at least 300 seconds
be set in the titration method for a determination using the oven
method.
The 885 Compact Oven SC is not equipped with a serial connec-
tion (RS232). Therefore, on the coulometer, switch the setting
Oven: to no under PARAM
▶ Preselect.
2 Deactivate the autostart function of the titrator.
8.1.2
Preparing the 885 Compact Oven SC
Selecting a method
Proceed as follows:
1 Switch on the instrument.
2 Use the arrow keys to select the line Method and press [OK].
3 Select the desired method and press [OK] again.
Inserting the conditioning vessel
1 Seal an empty sample vessel tightly with a septum seal.
2 Place the sealed vessel at the position Cond. Pos on the rack.