beautypg.com

2 preparing the 885 compact oven sc, Preparing the 885 compact oven sc – Metrohm 885 Compact Oven SC User Manual

Page 65

background image

■■■■■■■■■■■■■■■■■■■■■■

8 Carrying out a determination

885 Compact Oven SC

■■■■■■■■

55

It is advantageous for air bubbles to be stirred into the working
medium and forcefully broken up in the process. The air in the KF
titration cell must also be dried.

Creating a sample table at the titrator

1 Create a sample table on the KF titrator. For each blank, insert a line

which calls up a method for the determination of a blank.

You will find the necessary settings for the method in the KF titrator
manual.

Note

It is imperative that an Extraction time of at least 300 seconds
be set in the titration method for a determination using the oven
method.

The 885 Compact Oven SC is not equipped with a serial connec-
tion (RS232). Therefore, on the coulometer, switch the setting
Oven: to no under PARAM

Preselect.

2 Deactivate the autostart function of the titrator.

8.1.2

Preparing the 885 Compact Oven SC

Selecting a method

Proceed as follows:

1 Switch on the instrument.

2 Use the arrow keys to select the line Method and press [OK].

3 Select the desired method and press [OK] again.

Inserting the conditioning vessel

1 Seal an empty sample vessel tightly with a septum seal.

2 Place the sealed vessel at the position Cond. Pos on the rack.