5.4 Performing a sample series
■■■■■■■■■■■■■■■■■■■■■■
34
■■■■■■■■
885 Compact Oven SC
5.4
Performing a sample series
Blanks should always be at the first sample position so that the blank val-
ues determined are available with every sample processing.
5.4.1
Starting the sample series
Starting a sample series
1 Define the sample series
Press the [START] key.
You can now enter the quantity and the first rack position of the
samples to be processed.
2 Enter the number of samples
■
Select Number of samples and press [OK].
■
Enter the number of samples including the blanks.
■
Close the input dialog with [BACK] or Accept.
Note
Take care to ensure that the number of samples matches the num-
ber of sample data entered in the titrator.
3 Enter the rack position of the first sample
■
Select Next sample pos. and press [OK].
■
Enter the starting position of the sample series.
■
Close the input dialog with [BACK] or Accept.
The value for the number of samples remains saved for the next sam-
ple series. The position of the first sample is increased with each
method run.
You can still cancel the start of the sample series at this time with
[BACK] or [STOP].
4 Close the sample series dialog
Close the dialog with the [BACK] key.