7 Parameters

885 Compact Oven SC

■■■■■■■■

51

7 Parameters

Menu

Parameters

7.1

Automation

Temperature

Default value for the oven temperature.

Range

50 ... 250 °C (Increment: 1)

Default value

100 °C

Flow rate

Default value for the gas flow.

Range

10 ... 150 mL/min (Increment: 1)

Default value

50 mL/min

Gas supply

Selection of the gas supply.

Selection

pump | valve

Default value

pump

pump
Built-in air pump.
valve
Inlet valve for inert gas.

Gas type

Selection of the gas.

Selection

air | nitrogen | other

Default value

air

other
If other is selected, then a measuring factor for the gas that is used
must be specified. The measuring factor is used for the correct meas-
urement of the flow rate.

Measuring factor

Measuring factor for measuring the flow rate of the gas. This setting is
only possible if other is selected as the gas. The corresponding measuring
factor is automatically applied for air and nitrogen.