beautypg.com

4 lift settings (lift), Lift settings (lift) – Metrohm 885 Compact Oven SC User Manual

Page 56

background image

6.4 Lift settings (Lift)

■■■■■■■■■■■■■■■■■■■■■■

46

■■■■■■■■

885 Compact Oven SC

Program version

Version number of the oven modules' firmware. It cannot be modified.

6.4

Lift settings (Lift)

Menu

System Lift

Work position

The working height of the lift (sample position) can be set to the desired
value. This is accomplished by means of the direct operation of the lift. In
order to set this lift height, the sample rack must first be moved to a ran-
dom sample position (not the conditioning position).

Three functions can be selected from the function bar with [

] and []

and then executed by pressing [OK]:

Work pos. moves the lift to the current working height.

Up moves the lift 6 mm upward.

Down moves the lift 6 mm downward.

When this dialog page is exited, the respective current lift position will be
applied as Work position.

Note

Recommended settings:

With 6.2049.050 needle holder: 78 mm (do not move lower)

With 6.2049.040 needle holder: 96 mm

Range

0 ... 96 mm (Increment: 6)

Default value

78 mm

Conditioning position

The working height of the lift (at the conditioning position) can be set to
the desired value. This is accomplished by means of the direct operation of
the lift. In order to set this lift height, the sample rack must first be moved
with manual operation to the conditioning position of the rack.