Metrohm 798 MPT Titrino User Manual
Page 99
3.2 Remote control commands
798 MPT Titrino
95
"Setup",
continuation
├
.AutoInfo
Automatic message for changes
3.2.2.125.
│ ├
.Status Switch
AutoInfo on/off
ON, OFF
ditto
│ ├
.P
When mains is switched on
ON, OFF
ditto
│ ├
.T
Titrator infos
│ │ ├
.R
When “ready”
ON, OFF
ditto
│ │ ├
.G
When method started
ON, OFF
ditto
│ │ ├
.GC
When start is initiated
ON, OFF
ditto
│ │ ├
.S
When stopped
ON, OFF
ditto
│ │ ├
.B
Begin of method
ON, OFF
ditto
│ │ ├
.F
End of process
ON, OFF
ditto
│ │ ├
.E Error
ON,
OFF
ditto
│ │ ├
.H
When “hold"
ON, OFF
ditto
│ │ ├
.C
Continue after "hold" ON,
OFF
ditto
│ │ ├
.O
Conditioning OK
ON, OFF
ditto
│ │ ├
.N
Conditioning not OK
ON, OFF
ditto
│ │ ├
.Re
Request after start
ON, OFF
ditto
│ │ ├
.Si
Silo empty
ON, OFF
ditto
│ │ ├
.M
Entry in measuring point list
ON, OFF
ditto
│ │ ├
.EP
Entry in EP list
ON, OFF
ditto
│ │ ├
.RC
Recalculation of results done
ON, OFF
ditto
│ ├
.C
Comport infos
│ │ ├
.B1
When COM1 sends a report
ON, OFF
ditto
│ │ ├
.R1
When COM1 is ready again
ON, OFF
ditto
│ │ ├
.B2
When COM2 sends a report
ON, OFF
ditto
│ │ ├
.R2
When COM2 is ready again
ON, OFF
ditto
│ ├
.I
Changing an I/O input
ON, OFF
ditto
│ ├
.O
Changing an I/O output
ON, OFF
ditto
│
├
.Graphics
Changing the curve output
│ ├
.Grid
Grid on curve
ON, OFF
3.2.2.126.
│ ├
.Frame
Frame on curve
ON, OFF
ditto
│ ├
.Scale
Type of depending axis
Full, Auto
ditto
│ ├
.Recorder
Length of axes
│ │ ├
.Right
Length of meas value axis
0.2...0.5...1.00
ditto
│ │ ├
.Feed
Length of paper drive axis
0.01...0.05...1.00
ditto
│
├
.PowerOn
RESET (power on)
$G
3.2.2.127.
├
.Initialise
Set default values
$G
3.2.2.128.
│ ├
.Select
Selection of branch
ActMeth,Config,Silo,Calib
│
Assembly,Setup,All
ditto
├
.RamInit Initialization
of
working mem.
$G
3.2.2.129.
├
.InstrNo Device
Identification $G
3.2.2.130.
├
.Value
Input of device identification
8 ASCII characters
ditto