3.1 General rules
798 MPT Titrino
75
E130
Wrong sample. For SET with preset titration direction the first
measured value lies behind the endpoint.
Exit: The error message disappears on next startup.
E131
No EP set for SET.
Exit: The error message disappears on next startup.
E132
Silo empty and it has been started with open silo or empty silo has
been opened.
Exit: Send a silo entry.
E133
Silo full.
Exit: Send new command.
E134
No method. A method, which is required from the silo memory or in TIP, does
not exist.
Exit: The error message disappears on next startup.
E135
Check temp.sensor in MEAS T or with activated temperature monitoring.
Exit: Correct error or &m $S.
E136
Same buffer in CAL. Measured value of the second buffer differs less than 6
mV from the measured value of the first buffer.
Exit: Correct error or &m $S.
E137
XXX Bytes are missing so that the method, the silo line could not be stored or
not enough RAM for running TIP.
Exit: Send new command.
E155
No new silo result (C24 or C25).
Exit: The error message disappears on next start or on recalculation.
E157
No sequence defined in TIP.
Exit. The error message disappears on next start.
E158
A second TIP has been called up in TIP.
Exit: The error message disappears on next start.
E160
No new temporary variable.
Exit: The error message disappears on next start.
E161
Measurement range of the secondary measured value (temperature)
exceeded. The primary measured value (pH, U, Ipol, Upol) can also be
unstable.
Exit: Rectify error or &m $S.
E166
Save lines is "OFF" although a submethod of TIP includes an assignment to
C24 or C25.
Exit: The error message disappears on next start. Attention: The data of this
sample will not be stored.
E172
In TIP, a QuickMeas was started, without defining a measuring quantity.
Exit: The error message disappears on next start or &Mode.QuickMeas $S.
E177
Accessing to the memory card, the card was not (properly) inserted.
Exit. The error message disappears on next start.
E178
The date of changing the battery of the card is expired.
Exit. The error message disappears on next start.
E180
Memory card write-protected.
Exit: Send new command.
E181
Memory card not formatted.
Exit: Send new command.
E182
Memory card not accessible.
Exit: Send new command.
E183
A directory with the same name exists already on the memory card.
Exit: Send new command.