Metrohm 798 MPT Titrino User Manual
Page 122
3.2. Remote control commands
798 MPT Titrino
118
.MeasPt.Z1"6.34" Measured value (SET, MEAS) of the MPList, format de-
pends on the measured quantity
.MeasPt.Z2"25.8" Temperature of the MPList in °C
.EP.X"1.234" Volume coordinate of the EP
.EP.Y"5.34" Measured value coordinate of the EP
3.2.2.106. Info.ActualInfo.Display.L1
up to 32 ASCII characters
Info.ActualInfo.Display.L8
up to 32 ASCII characters
Info.ActualInfo.Display.DelAll $G
Lines of the display. The display can be written to from the computer. Pro-
ceed as follows:
1. Lock the display, see 3.2.2.119.
2. Delete the whole display (.DelAll).
3. For writing onto the display, the standard character set will be used.
4. Unlock the display, see 3.2.2.119
5. Delete the whole display (.DelAll).
6. Send a value to nod &Config.Aux.ResDisplay (see 3.2.2.78) to refresh the
display.
$Q sends the contents of the corresponding display line.
3.2.2.107. Info.ActualInfo.Comport.Number read
only
$Q sends the comport number of the Titrino where the PC is connected.
3.2.2.108. Info.Assembly.CycleTime read
only
Info.Assembly.ExV read
only
Inquiries regarding basic variables of the assembly: Cycle time in s, volume of
the active Exchange Unit in mL.
3.2.2.109. Assembly.Bur.Rates.Forward.Selected digital, analog
Assembly.Bur.Rates.Forward.Digital 0...150,
max.
Assembly.Bur.Rates.Reverse.Selected digital, analog
Assembly.Bur.Rates.Reverse.Digital 0...150,
max.
Expel and aspirating rate.
Digital or analog control. With digital control, the inputted value applies (in
mL/min). "max." means maximum possible rate with the Exchange Unit in
current use.
Analog means rate control with the analog potentiometer on Titrino.
3.2.2.110. Assembly.Bur.Fill
$G, $H, $C
$G starts the 'FILL' mode of the buret function.
3.2.2.111. Assembly.Bur.ModeDis
$G, $S, $H, $C
Assembly.Bur.ModeDis.Selected volume, time
Assembly.Bur.ModeDis.V 0.0001...0.1...9999
Assembly.Bur.ModeDis.Time 0.25...1...86400
Assembly.Bur.ModeDis.VStop 0.0001...9999, OFF
Assembly.Bur.ModeDis.AutoFill ON, OFF