3.2 Remote control commands
798 MPT Titrino
117
3.2.2.103. Info.ActualInfo.Assembly.Meas read
only
$Q sends the current measured value from the assembly.
3.2.2.104. Info.ActualInfo.Titrator.CyclNo read
only
Info.ActualInfo.Titrator.V read
only
Info.ActualInfo.Titrator.Meas read
only
Info.ActualInfo.Titrator.dVdt read
only
Info.ActualInfo.Titrator.dMeasdt read
only
Info.ActualInfo.Titrator.dMeasdV read
only
Info.ActualInfo.Titrator.ERC read
only
Info.ActualInfo.Titrator.T read
only
$Q sends the current values in the following formats:
DET
MET
SET
MEAS
CAL
CyclNo
127 127 127 127 127
V(ml)
1.2345 1.2345 1.2345 -
-
Meas:
pH
U, Ipol (mV)
Upol (uA)
T (°C)
3.345
-345.6
-12.5
-
3.345
-345.6
-12.5
-
3.6(mV)
-345.6
-12.5
-
3.345
-345.6
-12.5
25.0
3.345
-
-
-
dVdt
(ul/s) - - 2.5142
- -
dMeasdt
pH,U,Ipol mV/s
Upol (uA/s)
T (°C/s)
0.7957
0.7957
-
0.7957
0.7957
-
0.7957
0.7957
-
0.7957
0.7957
0.7957
0.7957
-
-
dMeasdV
(mV/ul) - - 10.6326
- -
ERC
34
- - - -
T
(°C)
25.9 25.9 -
25.9 -
NV: Not Valid. If in the signal drift is OFF in modes MEAS and CAL, the signal
drift is NV.
OV will be sent for "overrange".
A time frame can be set up from the cycle number and the cycle time (see
3.2.2.108). The cycle number is set to 0 at the start of a method and it is in-
cremented until the end of the method.
3.2.2.105. Info.ActualInfo.MeasPt.Index read
only
Info.ActualInfo.MeasPt.X read
only
Info.ActualInfo.MeasPt.Y read
only
Info.ActualInfo.MeasPt.Z1 read
only
Info.ActualInfo.MeasPt.Z2 read
only
Info.ActualInfo.EP.Index read
only
Info.ActualInfo.EP.X read
only
Info.ActualInfo.EP.Y read
only
$Q sends the last entry into the measuring point list (.MeasPt) or the last entry
into the list of EP's with DET, MET, SET.
.MeasPt.X"165" Volume (DET, MET), time (SET), resp. of the MPList in s
.MeasPt.Y"3.654" Measured value (DET, MET), volume (SET), resp. of the
MPList in mL