3.2. Remote control commands
798 MPT Titrino
108
Format the memory card. The action is carried out if "$G" is transmitted to the
corresponding node.
Do not use blank characters before and after name!
3.2.2.60. MemoryCard.FreeMem read
only
Free memory on the card. $Q sends number of free bytes, e.g. "4928".
3.2.2.61. MemoryCard.BatteryChange $G
MemoryCard.BatteryChange.Date YYYY-MM-DD
Date for changing battery. The date will be set with "$G".
3.2.2.62. MemoryCard.List.Card.1.Name read
only
MemoryCard.List.Card.1.Bytes read
only
for each directory
List of all directories on the memory card with the following information:
.Name:
Name of the directory
.Bytes:
Number of bytes used by the directory
3.2.2.63. MemoryCard.List.ActDir.1.Name read
only
MemoryCard.List.ActDir.1.Mode read
only
MemoryCard.List.ActDir.1.Quantity read
only
MemoryCard.List.ActDir.1.Bytes read
only
MemoryCard.List.ActDir.1.Checksum read
only
for each method
List of all methods of the current card directory with the following information:
.Name:
Name of the method
.Mode: Mode
.Quantity: Measured
quantity
.Bytes:
Number of bytes used by the method
.Checksum: Checksum of the method, see 3.2.2.94.
3.2.2.64. Config.Monitoring.Validation.Status ON,
OFF
Config.Monitoring.Validation.Interval 1...365...9999
Config.Monitoring.Validation.Counter
0...9999
Config.Monitoring.Validation.ClearCount $G
Monitoring of validation.
.Interval: Time interval in days for validation.
.Counter: Time counter in days since last validation.
.ClearCount: Clears the above counter.
3.2.2.65. Config.Monitoring.Calibration.Status ON,
OFF
Config.Monitoring.Calibration.MeasInput
1, 2, diff
Config.Monitoring.Calibration.Interval 1...7...9999
Config.Monitoring.Calibration.Counter
0...9999
Monitoring of pH calibration.
.MeasInput:Measuring input.
.Interval: Time interval in days for calibration from the last calibration date.