.MinRate: Minimum titration rate in ul/min.
3.2.2.30. Mode.Parameter.SET1.Stop.Type drift, time
Mode.Parameter.SET1.Stop.Drift 1...20...999
Mode.Parameter.SET1.Stop.Time 0...10...999, inf
Mode.Parameter.SET1.Stop.StopT 0...99
999, OFF
Parameters for SET: Type and size of the stop criterion of the titration.
.Type: Type of stop criterion after stop drift or switch-off delay time.
.Drift: Stop drift in ul/min. Applies when "drift" has been selected.
.Time: Switch-off delay time in s. Applies when "time" has been selected. "inf"
means infinite.
.StopT: Stop time in s. Applies when "time" has been selected and the value of
.Time is set to "inf.".
3.2.2.31. Mode.Parameter.TitrPara.Direction +,
-, auto
Parameters for SET: Titration direction.
"auto" means the titration direction is determined automatically by the instru-
ment. If 2 EP's have been set in a SET titration, the titration direction is given
by the two EP's. The entry of the titration direction is then invalid.
3.2.2.32. Mode.Parameter.TitrPara.XPause
0...999 999
Parameter for SET: Pause time in s. Runs before dosing the start volume.
3.2.2.33. Mode.Parameter.TitrPara.ExtrT
0...999 999
Parameter for SET: Extraction time in s.
3.2.2.34. Mode.Parameter.TitrPara.TDelta
1...2...999 999
Parameter for SET: Time interval in s for the entry of a measurement point in
the list of measured points.
3.2.2.35. Mode.Parameter.Presel.Cond ON, OFF
Mode.Parameter.Presel.DriftDisp ON, OFF
Mode.Parameter.Presel.DCor.Type auto,
man., OFF
Mode.Parameter.Presel.DCor.Value 0.0...99.9
Parameters for SET:
.Cond: Conditioning ON/OFF
.DriftDisp: Drift display during conditioning ON/OFF.
.DCor.Type: Type of drift take-over for the drift correction. auto: Take-over of
the drift value at start.
.DCor.Value: Drift value for the manual drift correction.
3.2.2.36. Mode.Parameter.Measuring.SignalDrift
pH, U, Ipol, T:
0.5...999, OFF
Upol: 0.05...99.9, OFF
Mode.Parameter.Measuring.UnitSigDrift read
only
Mode.Parameter.Measuring.EquTime 0...9999, OFF