Info – Metrohm 798 MPT Titrino User Manual
Page 94
3.2. Remote control commands
798 MPT Titrino
90
&Info
Object Description
Input
range
Reference
& Root
:
├
Info
Current data
:
├
.Report
Transmission of formatted reports $G 3.2.2.92.
│ ├
.Select
Report type
configuration, parameters,
│
smpl data, statistics, silo, calib
│
C-fmla, def, user method, full,
│
short, mplist, curve, derive, comb,
│
adj para,
scalc full, scalc srt, calc,
│
act dir, mem card, all, ff
ditto
│
├
.CalibrationData
pH calibration data
$G 3.2.2.93.
│ ├
.Inp1
For measuring input 1
│
│ ├
.pHas Asymmetry
pH
0...7.00...±20.00 ditto
│
│ ├
.Slope
Slope of electrode
0...1.000...±9.999 ditto
│
│ ├
.Temp Calibration
temperature
-170.0...25.0...500.0 ditto
│
│ ├
.Date
Date of calibration
read only
ditto
│
│ ├
.ElectrodeId
Id of the calibrated electrode
read only
ditto
│ ├
.Inp2
For measuring input 2, as for input 1
│ ├
.Diff
For differential input, as for input 1
│
├
.Checksums
Checksums
$G 3.2.2.94.
│ ├
.MPList
Checksum of meas.point list
read only
ditto
│ ├
.ActualMethod
Checksum of current method
read only
ditto
│
├
.DetermData
Determination data
$G 3.2.2.95.
│ ├
.Write
Read/write for several nods
ON, OFF
│ ├
.ExV
Volume of Exchange/Dosing unit read
only/read+write
ditto
│ ├
.MPList
Measuring point list
│
│ ├
.1
Measuring point 1
│
│
│ ├
.Attribute Attribute
read
only/read+write
ditto
│
│
│ ├
.X
X coordinate
read only/read+write
ditto
│
│
│ ├
.Y
Y coordinate
read only/read+write
ditto
│
│
│ ├
.Z1
Z1 coordinate
read only/read+write
ditto
│
│
│ ├
.Z2
Z2 coordinate
read only/read+write
ditto
│
│
│
for each measuring point
│
├
.TitrResults
Titration results
│ ├
.RS
Calculated results
│ │ ├
.1
1
st
result
│ │ │ ├
.Value Value
read
only
3.2.2.96.
│ │ ├
up to 9 results
"Info",
continuation
│ ├
.EP
Endpoint
│ │ ├
.1
1
st
result
│
│ │ ├
.V
Value
read only
ditto
│
│ │ ├
.Meas
Measured value
read only
│
│ │ ├
.Mark
Mark if more than 1 EP per window read only
│ │ ├
up to 9 EP's
│ ├
.Var
Variables C4X
│ │ ├
.C40
Start measured value
read only/read+write
ditto
│ │ ├
.C41
Titration end volume
read only/read+write
│ │ ├
.C42
Titration time
read only/read+write