3.2. Remote control commands
798 MPT Titrino
102
window: Only EP's that lie within set windows are recognized.
OFF: The EP evaluation is switched off.
.Recognition.Window.1.LowLim: Lower limit for window in pH, mV (with U
and Ipol) or µA (with Upol).
.Recognition.Window.1.UpLim: Upper limit for window in pH, mV (with U and
Ipol) or µA (with Upol).
Windows are opened until the lower limit is set to OFF. For every
expected EP, an individual window must be set, see page 22.
.FixEP.1.Val: Fix-EP's in pH, mV (for U, Ipol) resp. µA (for Upol). Fix EP's are
evaluated until the setting OFF is found.
.pK: pK or HNP evaluation. Possible only in pH and U titrations.
3.2.2.25. Mode.Parameter.Presel.IReq
id1, id1&2, all, OFF
Mode.Parameter.Presel.SReq
value, unit, all, OFF
Parameters for DET, MET, SET, MEAS: Automatic inquiry after the start of the
determination. From such an inquiry, the determination continues if the re-
quested entry/entries is/are made, e.g. &SmplData.OFFSilo.Id1 (see 3.2.2.84)
or with &M $G, see 3.2.2.1.
$H is not possible in requests.
3.2.2.26. Mode.Parameter.Presel.LimSmplSize.Status ON,
OFF
Mode.Parameter.Presel.LimSmplSize.LoLim
0.0...999 999
Mode.Parameter.Presel.LimSmplSize.UpLim 0.0...999 999
Limit control for the sample size.
3.2.2.27. Mode.Parameter.Presel.ActPuls ON, OFF
for SET:
first, all, cond., OFF
Output of a pulse on the I/O line "Activate", see page 167.
3.2.2.28. Mode.Parameter.SET1.EP pH:
0...±20.00, OFF
U,
Ipol:
0...±2000, OFF
Upol:
0...±200.0, OFF
Mode.Parameter.SET1.UnitEp read
only
Parameters for SET: Setting the 1st endpoint as pH value, in mV (with U and
Ipol) resp. µA (with Upol). The corresponding unit can be read with .UnitEP.
If the value is on "OFF", no further nodes will appear from SET1.
3.2.2.29. Mode.Parameter.SET1.Dyn pH:
0.01...20.00, OFF
U,
Ipol:
1...2000, OFF
Upol:
0.1...200.0, OFF
Mode.Parameter.SET1.UnitDyn read
only
Mode.Parameter.SET1.MaxRate 0.01...10...150, max.
Mode.Parameter.SET1.MinRate 0.01...25.0...999.9
Parameters for SET: Control parameters, see page 30.
.Dyn:
Dynamics, control range in pH, mV (with U and Ipol) or µA (with
Upol). The corresponding unit can be read with .UnitDyn.
.MaxRate: Maximum allowed titration rate in mL/min. Max. means maximum
possible rate with the Exchange Unit in current use.