3.2 Remote control commands
798 MPT Titrino
111
$G sets all RS settings. The changes are performed only if the instrument is
inactive. After the setting of the interface parameters, wait at least 2 s to allow
the components to equilibrate.
Settings of the values for the data transmission via the RS interface: baud
rate, data bit, stop bit, parity and type of handshake, see also page 126ff.
Baud rates >9600 need a PC which is equipped accordingly (e.g. with 16550
component).
The setting of the values must be initiated with $G immediately after entry of
the values.
3.2.2.82. Config.ComVar.C30
with up to .C39, etc.
0... +999 999
Values of the common variables from C30 up to C39. Insert the common
variables directly or describe the determination results directly from the
method, see 3.2.2.7
3.2.2.83. SmplData.Status ON, OFF
On/off switching of silo memory. When the silo memory is switched on, the
sample data are fetched from the lowest valid silo line.
3.2.2.84. SmplData.OFFSilo.Id1
up to 8 ASCII characters
SmplData.OFFSilo.Id2
up to 8 ASCII characters
SmplData.OFFSilo.Id3
up to 8 ASCII characters
SmplData.OFFSilo.ValSmpl
6-digits, sign and decimal point
SmplData.OFFSilo.UnitSmpl
up to 5 ASCII characters
Current sample data.
The identifications Id1...Id3 can be used in formulas as sample-specific calcu-
lation constants C21...C23.
If "no unit" is desired for the unit of the sample size, the blank string must be
entered.
3.2.2.85. SmplData.ONSilo.Counter.MaxLines read
only
SmplData.ONSilo.Counter.FirstLine read
only
SmplData.ONSilo.Counter.LastLine read
only
Information on silo memory.
.MaxLines: Maximum possible number of silo lines.
.FirstLine: Lowest valid silo line.
.LastLine: Last occupied silo line.
3.2.2.86. SmplData.ONSilo.EditLine.1.Method
up to 8 ASCII characters
SmplData.ONSilo.EditLine.1.Id1
up to 8 ASCII characters
SmplData.ONSilo.EditLine.1.Id2
up to 8 ASCII characters
SmplData.ONSilo.EditLine.1.Id3
up to 8 ASCII characters
SmplData.ONSilo.EditLine.1.ValSmpl 6-digits, sign and dec.point
SmplData.ONSilo.EditLine.1.UnitSmpl
up to 5 ASCII characters
SmplData.ONSilo.EditLine.1.C24 read
only
SmplData.ONSilo.EditLine.1.C25 read
only
SmplData.ONSilo.EditLine.1.Mark read
only
etc., up to .255