beautypg.com

Metrohm 798 MPT Titrino User Manual

Page 118

background image

3.2. Remote control commands

798 MPT Titrino

114

The checksums can be used to identify the content of a file unequivocally,

e.g. files with identical content

have identical results of the checksums. An empty file has checksum "0". The

calculation of the checksums is triggered with $G.

.MPList: Result of the checksum of the current measuring point list.

.ActualMethod: Result of the checksum of the current method in the working

memory. Identical methods with different method names have the

same results of the checksum.

3.2.2.95. Info.DetermData $G

Info.DetermData.Write ON,

OFF

Info.DetermData.MPList.1.Attribute read

only/read+write

Info.DetermData.ExV read

only/read+write

Info.DetermData.MPList.1.X read

only/read+write

Info.DetermData.MPList.1.Y read

only/read+write

Info.DetermData.MPList.1.Z1 read

only/read+write

Info.DetermData.MPList.1.Z2 read

only/read+write

for every measuring point

Determination data in hexadecimal format. A measuring point list is available

in mode DET, MET, SET, and MEAS.

Recalculation of the measuring data is triggered with $G.

.Write:

With "ON", the following nodes can be overwritten:

&Info.DetermData.MP.List, &Info.TitrResults.Var.C4X (X = 0...5),

&Info.TitrResults.TempVar.C7X (X = 0...9), and &Mode.Name.

.ExV:

Volume of the exchange unit, with which the determination was

executed

.MPList.1.Attribute: Attribute

.MPList.X: X coordinate, time

.MPList.Y: Y coordinate, volume

.MPList.Z1: Z1 coordinate, measuring value

.MPList.Z2: Z2 coordinate, temperature

3.2.2.96. Info.TitrResults.RS.1.Value read

only

etc., up to .9

Info.TitrResults.EP.1.V read

only

Info.TitrResults.EP.1.Meas read

only

etc., up to .2

Info.TitrResults.Var.C40 read

only/read+write

etc., up to .C47

Info.TitrResults.Var.DTime read

only/read+write

.RS: Values of the calculated results.

.EP: Endpoints with DET, MET, SET:

Volume coordinate in mL, e.g. "1.2340"

Measured value coordinate in pH "5.12", mV (with U and Ipol) "-

241" or µA (with Upol) "43.7".

.Var: Various variables. You may overwrite the variables C40...C45, see

3.2.2.95.

C40: Initial measured value in pH "5.12", mV (with U and Ipol)

"41", µA (with Upol) "43.7" or °C (with T) "25.0". In MEAS final

measured value.

C41: End volume with SET in ml, "12.5360".