3.2 Remote control commands
798 MPT Titrino
119
Dispensing mode with parameters. The dispensing mode can only be started
and stopped via the RS Control. During a running dosification, no method can
be started at the Titrino.
.Selected: Dispensing of volume increments or during a preset time.
.Volume, .Time: Size of the volume increments or entry of time.
.VStop:
Limit volume for the dispensing.
.AutoFill: ON means automatic filling after every dispensing.
3.2.2.112. Assembly.Meas.Status ON, OFF
Assembly.Meas.MeasInput
1, 2, Diff., Ipol, Upol, Temp
Assembly.Meas.Ipol ±127...1...+127
Assembly.Meas.Upol ±1270...400...+1270
Measurement in assembly. The measuring function can only be started via RS
Control. When the measuring function is switched on, no method can be
started at the Titrino.
.Input:
Selection of the potentiometric measuring input 1, 2, diff., polar-
ized electrodes or temperature.
.Ipol:
Polarization current in µA.
.Upol:
Polarization potential in mV, entry in steps of 10 mV.
3.2.2.113. Assembly.Outputs.AutoEOD ON, OFF
Assembly.Outputs.SetLines $G
Assembly.Outputs.SetLines.L0
active, inactive, pulse, OFF
up to .L13
Assembly.Outputs.ResetLines $G
Setting the I/O output lines.
.AutoEOD: The automatic output of the EOD (End of Determination) at the
end of the determination can be switched off. Thus, for example,
in conjunction with a Titrino several determinations can be per-
formed in the same beaker. Before AutoEOD is switched on, line 3
must be set to "OFF".
.SetLines: With $G, all lines are set.
.SetLines.LX: Set the line LX. "active" means setting of a static signal, "inac-
tive" means resetting of the signal, "pulse" means output of a
pulse of app. 150 ms, "OFF" means the line is not operated, see
also page 167.
Warnings:
• If you have "AutoEOD" to "ON", an active line 3 is set to "inac-
tive" by the EOD pulse.
• L6 is the line of the activate pulse. An active line 6 is set to "in-
active" by the activate pulse.
• L5 is the error line. It is continuously controlled by the Titrino
program and can therefore not be set freely.
Line assignments in Titrino program:
L0
Ready, inactive state
L1 Conditioning
OK
L2
Titration in progress
L3
EOD (End Of Determination)
L4
Can be set in TIP
L5 Error