8.8 Remote control via an RS-232 connection
■■■■■■■■■■■■■■■■■■■■■■
72
■■■■■■■■
865 Dosimat plus
Command
Function
Comment
$D
Scan instrument sta-
tus
Answers: Ready;0 , Busy;0 , or Hold;0 (0
= no message)
If a message on the instrument requires
the interaction of the user, the answer of
the status scan displays the corresponding
message number. Example:
Busy;010-119
= "Check buret unit"
The message can be quit with [OK] or
[Cancel], see below.
$A
Quit message
Confirm the message on the instrument
with [OK]
A mandatory status scan providing the
message number must take place immedi-
ately before quitting the message, see
above.
$A(OK), $A(CANCEL)
Quit message
Confirm the message with [OK] or [Can-
cel]
$A(YES), $A(NO)
Quit message
Confirm the message with [Yes] or [No]
$L(method name)
Load method
The method name has to be known and
unique.
$Q(variable)
Request variable value
List of variables, see below
Variable
Comment
VOLUME
Dosed volume
TITER
Titer of selected solution
CONC
Concentration of selected solution
C00
Sample size (DOS only)
RESULT
Result (DOS only)
RATE
Dosing rate (XDOS only)
TIME
Dosing time (XDOS only)
CONTENT
Content (CNT only)
The values of the variables are only available after the end of a determina-
tion (in the status 'ready').