6 Parameters and dosing modes
865 Dosimat plus
■■■■■■■■
47
6.2.4
Dosing parameters
Menu
▶ Parameters ▶ Dosing parameters
Dosing criteria
Selection of the preset dosing criteria.
Selection
Rate/Time | Volume/Rate | Volume/Time
Default value
Volume/Rate
Rate/Time
Dosing is carried out at the selected dosing rate for the amount of time
specified. The time is also running while the cylinder is being filled.
Volume/Rate
The volume specified is dosed at the selected dosing rate.
Volume/Time
The volume specified is dosed for the time period selected. The dosing
rate necessary for this, is calculated according to these specifications.
The time needed for filling the dosing cylinder and for switching over
the flat stopcock is taken into account thereby.
Volume
The volume to be dosed.
Range
0.0000 ... 99999.9 mL
Default value
10.0000 mL
Dosing rate
The rate at which dosing takes place. The maximum dosing rate depends
on the cylinder volume (see Chapter 8.1.1, page 61).
Range
0.01 ... 150.00 mL/min
Selection
max.
Default value
max.
Filling rate
Rate at which the dosing cylinder is to be refilled. The maximum filling
rate depends on the cylinder volume (see Chapter 8.1.1, page 61).
Range
0.01 ... 150.00 mL/min
Selection
max.
Default value
max.
Volume limit
Safety limit for limiting the maximum volume to be dosed.
Range
0.01 ... 99999.9 mL