3 tubing connections, Tubing connections – Metrohm 865 Dosimat plus User Manual
Page 14
1.4 Safety instructions
■■■■■■■■■■■■■■■■■■■■■■
4
■■■■■■■■
865 Dosimat plus
Warning
Never open the housing of the instrument. The instrument could be
damaged by this. There is also a risk of serious injury if live components
are touched.
There are no parts inside the housing which can be serviced or replaced
by the user.
Mains voltage
Warning
An incorrect mains voltage can damage the instrument.
Only operate this instrument with a mains voltage specified for it (see
rear panel of the instrument).
Protection against electrostatic charges
Warning
Electronic components are sensitive to electrostatic charges and can be
destroyed by discharges.
Always pull the mains cable out of the mains connection socket before
connecting or disconnecting electrical appliances on the rear panel of
the instrument.
1.4.3
Tubing connections
Caution
Leaks in tubing connections are a safety risk. Tighten all connections
well by hand. Avoid applying excessive force to tubing connections.
Damaged tubing ends lead to leakage. Appropriate tools can be used
to loosen connections.
Check the connections regularly for leaks.