2 selecting a solution, 3 definition of the target solution, Selecting a solution – Metrohm 865 Dosimat plus User Manual
Page 61: Definition of the target solution

■■■■■■■■■■■■■■■■■■■■■■
6 Parameters and dosing modes
865 Dosimat plus
■■■■■■■■
51
6.3.2
Selecting a solution
Menu
▶ Parameters
Solution
Selection of the solution from the solution list. We recommend always
selecting the solution. Solutions are defined under System
▶ Solutions.
For exchange units with integrated data chip, a check is made in the
method run to verify whether the correct solution has been attached. For
exchange units without integrated data chip, only the cylinder volume is
checked. The validity of the titer is checked for the selected solution at the
start of the determination.
Selection
Selection of configured solutions | not defined
Default value
not defined
not defined
No check takes place.
6.3.3
Definition of the target solution
Menu
▶ Parameters ▶ Content definitions
Content
Target concentration of the solution.
Input
10 characters
Range
0.000000001 ... 9999999999
Default value
1.0
Content unit
Unit of the target concentration.
The unit of the target concentration determines the calculation formula
for determining the volume to be dosed.
Selection
% | g/L | mg/L | mmol/L | mmol/kg | mol/L |
mol/kg | ppm
Default value
mol/L
Molar mass
Molar mass of the starting substance.
If selected unit is: mmol/L, mmol/kg, mol/L or mol/kg.
Input
10 digits
Range
0.000000001 ... 9999999999 g/mol
Default value
1.0 g/mol