Metrohm 865 Dosimat plus User Manual
Page 40

5.1 Basic settings
■■■■■■■■■■■■■■■■■■■■■■
30
■■■■■■■■
865 Dosimat plus
Contrast
The contrast of the display can be adjusted with the arrow keys [
⇦] and
[
⇨].
■
[
⇦]: the contrast will be decreased by one step each time the key is
pressed.
■
[
⇨]: the contrast will be increased by one step each time the key is
pressed.
Range
150 ... 240
Default value
212
Note
Alternatively, the contrast can also be modified in the following man-
ner:
Keep the red [STOP] key pressed down. As soon as the progress bar
appears, also press the arrow key [
⇩] or [⇧] repeatedly.
This method will however cause the contrast to be modified by several
steps.
Beep
If this parameter is activated, then a short beep will be heard in the fol-
lowing cases:
■
When a key is pressed.
■
At the end of the determination.
Selection
on | off
Default value
on
PREP warning
If this parameter is activated, then the recommendation to carry out the
PREP (preparing) function will be displayed in the following cases:
■
After the instrument has been switched on.
■
Each time a buret unit has been attached.
This function causes all tubings and the cylinder to be rinsed (see Chapter
8.1.2, page 61).
Selection
on | off
Default value
on