2 selecting a solution, 3 selecting the operating mode, Selecting a solution – Metrohm 865 Dosimat plus User Manual
Page 55: Selecting the operating mode

■■■■■■■■■■■■■■■■■■■■■■
6 Parameters and dosing modes
865 Dosimat plus
■■■■■■■■
45
time of the output of the report. This requires that the name of the solu-
tion be specified under Menu
▶ Parameters.
Performing a dosing
Pressing the [GO] key causes the dosing to run automatically. Three dos-
ing criteria are applied for this process. The specified dosing criteria are
observed in all cases. The dosing criterion that is not defined in a specific
case (volume, time or dosing rate) is calculated by the instrument and
shown on the display. The dosing is optimized according to the specifica-
tions. Continuous dosing is ensured in tandem operation, if the dosing
and filling rates have been set accordingly.
The dosing can be canceled with the [FILL] key. If the Auto fill parame-
ter is switched on, then the dosing cylinder will be filled.
The dosing criteria are shown on the display. They can be deleted by
pressing the [FILL] key once again.
6.2.2
Selecting a solution
Menu
▶ Parameters
Solution
Selection of the solution from the solution list. We recommend always
selecting the solution. Solutions are defined under System
▶ Solutions.
For exchange units with integrated data chip, a check is made in the
method run to verify whether the correct solution has been attached. For
exchange units without integrated data chip, only the cylinder volume is
checked. The validity of the titer is checked for the selected solution at the
start of the determination.
Selection
Selection of configured solutions | not defined
Default value
not defined
not defined
No check takes place.
6.2.3
Selecting the operating mode
Menu
▶ Parameters
Buret setup
If continuous dosing without interruption is required, then the tandem
mode can be selected. An additional 805 Dosimat is required for this pur-
pose.
Selection
Single | Tandem
Default value
Single