2 selecting a solution, 3 definitions of the sample transfer, Selecting a solution – Metrohm 865 Dosimat plus User Manual
Page 66: Definitions of the sample transfer

6.4 Pipetting and diluting (LQT)
■■■■■■■■■■■■■■■■■■■■■■
56
■■■■■■■■
865 Dosimat plus
6.4.2
Selecting a solution
Menu
▶ Parameters
Solution
Selection of the solution from the solution list. We recommend always
selecting the solution. Solutions are defined under System
▶ Solutions.
For exchange units with integrated data chip, a check is made in the
method run to verify whether the correct solution has been attached. For
exchange units without integrated data chip, only the cylinder volume is
checked. The validity of the titer is checked for the selected solution at the
start of the determination.
Selection
Selection of configured solutions | not defined
Default value
not defined
not defined
No check takes place.
6.4.3
Definitions of the sample transfer
Menu
▶ Parameters ▶ Transfer definitions
Transfer volume
Sample volume to be transferred.
The sample volume selected must not be too large in size. The sum of the
transfer volume and the air bubbles must be less than 98 % of the cylinder
volume used. If this value is exceeded, then no precise sample transfer is
possible.
Range
0.0100 ... 49.5000 mL
Default value
1.0000 mL
Air gap
Volume of the air gap between sample and carrier liquid.
The air gap in the pipetting tubing should have a length of at least 5 mm.
It should however also not be too large in size.
Range
0.0100 ... 49.5000 mL
Default value
0.0500 mL
Safety air gap
Volume of an air bubble at the dosing tip after the aspiration of the sam-
ple. Do not define an air bubble too large in size.
Range
0.0000 ... 49.5000 mL
Default value
0.0000 mL