4 dosing parameters, 5 reports, Dosing parameters – Metrohm 865 Dosimat plus User Manual
Page 67: Reports
■■■■■■■■■■■■■■■■■■■■■■
6 Parameters and dosing modes
865 Dosimat plus
■■■■■■■■
57
Diluting volume
Volume of liquid for after rinsing or diluting the sample.
Range
0.0000 ... 9999.99 mL
Default value
0.0000 mL
6.4.4
Dosing parameters
Menu
▶ Parameters ▶ Dosing parameters
Dosing rate
The rate at which dosing takes place. The maximum dosing rate depends
on the cylinder volume (see Chapter 8.1.1, page 61).
Range
0.01 ... 150.00 mL/min
Selection
max.
Default value
max.
Filling rate
Rate at which the dosing cylinder is to be refilled. The maximum filling
rate depends on the cylinder volume (see Chapter 8.1.1, page 61).
Range
0.01 ... 150.00 mL/min
Selection
max.
Default value
max.
Aspiration rate
The rate at which the sample is aspirated. The maximum rate depends on
the cylinder volume (see Chapter 8.1.1, page 61).
Range
0.01 ... 150.00 mL/min
Selection
1.0
Default value
1.0
6.4.5
Reports
Menu
▶ Parameters ▶ Reports
The reports that will be printed out automatically in connection with a
determination are defined under Reports.
Results
The result report contains the calculated result and additional specifica-
tions.
Selection
off | on
Default value
off