2 selecting a solution, 3 dosing parameters, Selecting a solution – Metrohm 865 Dosimat plus User Manual
Page 50: Dosing parameters

6.1 Manual dosing (DOS)
■■■■■■■■■■■■■■■■■■■■■■
40
■■■■■■■■
865 Dosimat plus
The [FILL] key can be used at the end of a determination to initiate the
filling of the dosing cylinder. This automatically initiates the calculation of
the result and the printing out of a result report, if the appropriate settings
have been made; see following chapters.
The result of the calculation is shown under the volume display. The result
and the dosing volume can be deleted by pressing the [FILL] key once
more.
6.1.2
Selecting a solution
Menu
▶ Parameters
Solution
Selection of the solution from the solution list. We recommend always
selecting the solution. Solutions are defined under System
▶ Solutions.
For exchange units with integrated data chip, a check is made in the
method run to verify whether the correct solution has been attached. For
exchange units without integrated data chip, only the cylinder volume is
checked. The validity of the titer is checked for the selected solution at the
start of the determination.
Selection
Selection of configured solutions | not defined
Default value
not defined
not defined
No check takes place.
6.1.3
Dosing parameters
Menu
▶ Parameters ▶ Dosing parameters
Dosing rate
The rate at which dosing takes place. The maximum dosing rate depends
on the cylinder volume (see Chapter 8.1.1, page 61).
Range
0.01 ... 150.00 mL/min
Selection
max.
Default value
max.
Filling rate
Rate at which the dosing cylinder is to be refilled. The maximum filling
rate depends on the cylinder volume (see Chapter 8.1.1, page 61).
Range
0.01 ... 150.00 mL/min
Selection
max.
Default value
max.