Metrohm 815 Robotic Soliprep User Manual
Page 87

■■■■■■■■■■■■■■■■■■■■■■
3 Robotic Flexible Soliprep 2.815.2110
815 Robotic Soliprep
■■■■■■■■
79
12 Move the lift to the Target position = Work position. The Poly-
tron aggregate must sufficiently come into contact with the absorb-
ent material. Correct the lift position if necessary.
13 If the position has had to be corrected, switch to the Assign posi-
tion tab.
14 Select the Work position for = Special beaker 2 in the Lift posi-
tion field.
15 Click on the associated [Assign] button.
16 Switch back to the Move tab.
Lift positions for stripping needles and filters
1 In the manual control in the left-hand window, click on the Tower 1
item under 815_1 (815 Robotic …) and then select the Move tab.
2 Fasten a syringe needle or a membrane filter to the Luer adapter of
the transfer robotic arm.
3 Under Robotic arm position, select the Target position = Exter-
nal 1 and click on [Start].
4 Move the lift downward until the thick sleeve of the Luer adapter is
approx. 1 cm underneath the deflector. Under Lift position, click on
the arrow key [Arrow down] and also (if required) [Arrow up] in
order to set a suitable position.
The Luer adapter must be positioned at a distance of approx. 2 cm
from the deflector. Its thick sleeve must be positioned approx. 1 cm