11 connecting and setting up peristaltic pumps, Connecting and setting up peristaltic pumps – Metrohm 815 Robotic Soliprep User Manual
Page 32

2.1 Installation
■■■■■■■■■■■■■■■■■■■■■■
24
■■■■■■■■
815 Robotic Soliprep
2.1.11
Connecting and setting up peristaltic pumps
The two 772 Pump Unit peristaltic pumps are used for aspirating the titra-
tion vessels and the rinsing station.
772 Pump Unit for the rinsing station
6.1812.000
6.1812.000
6.1820.050
6.1826.100
1
2
3
4
1 Cut one piece of 6.1826.100 pump tubing down to a length of
approx. 17 cm and attach a 6.1820.050 tubing olive at both ends.
2 Fasten a piece of 6.1812.000 PTFE tubing leading to a waste con-
tainer to one of the tubing olives. Attach the 6.1812.000 PTFE tubing
that is connected to the rinsing station as a drain connection to the
other tubing olive.
3 Release the locking lever by rotating the clamping screw and open
the pressure clamp.
4 Loop the pump tubing around the rotor and fasten with the tubing
clamps. The rotor turns in clockwise direction. The inlet tubing must
therefore be attached below, the outlet tube above.
- 915 KF Ti-Touch (382 pages)
- 800 Dosino (53 pages)
- 767 Calibrated Reference (23 pages)
- 940 Professional IC Vario ONE/SeS/Prep 2 (54 pages)
- 754 Dialysis Unit (49 pages)
- 815 Robotic Soliprep for LC (76 pages)
- Vision Manual (207 pages)
- tiamo 2.1 Manual (1532 pages)
- 825 Lab Link (37 pages)
- 808 Titrando (70 pages)
- 902 Titrando (52 pages)
- 756 KF Coulometer (163 pages)
- 756 KF Coulometer (162 pages)
- 940 Professional IC Vario ONE/LPG (98 pages)
- 850 Professional IC Anion MCS Prep 3 (154 pages)
- 850 Professional IC Anion MCS Prep 3 (152 pages)
- 904 Titrando (58 pages)
- 850 Professional IC Anion MSM-HC MCS Prep 2 (150 pages)
- 930 Compact IC Flex Oven/ChS/Deg (47 pages)
- 872 Extension Module Liquid handling (64 pages)
- 814 USB Sample Processor (90 pages)
- 814 USB Sample Processor (91 pages)
- 940 Professional IC Vario (43 pages)
- Vision – Tutorial (40 pages)
- 799 GPT Titrino (242 pages)
- 889 IC Sample Center (68 pages)
- 761 Compact IC (228 pages)
- 851 Titrando (100 pages)
- 748 DH Sample Changer (32 pages)
- 940 Professional IC Vario ONE/SeS/HPG (51 pages)
- 896 Professional Detector – Amperometry (62 pages)
- 877 Titrino plus (139 pages)
- 881 Compact IC pro – Anion (129 pages)
- 940 Professional IC Vario ONE/ChS/HPG (112 pages)
- 930 Compact IC Flex Deg (41 pages)
- 840 PC Control 5.0 / Touch Control (351 pages)
- 940 Professional IC Vario ONE/Prep 1 (45 pages)
- 776 Dosimat (42 pages)
- 717 Sample Changer (36 pages)
- 815 Robotic USB Sample Processor XL (113 pages)
- 815 Robotic USB Sample Processor XL (114 pages)
- 940 Professional IC Vario ONE/SeS/PP (126 pages)
- 838 Advanced Sample Processor Installation Instructions (109 pages)
- 700 Dosino (55 pages)
- 719 S Titrino (152 pages)