Metrohm 815 Robotic Soliprep User Manual
Page 86

3.3 Setting lift positions
■■■■■■■■■■■■■■■■■■■■■■
78
■■■■■■■■
815 Robotic Soliprep
Lift positions for special beakers
1 Under Rack position, enter the Target position = Special
beaker 1 and click on [Start].
2 Place a sample beaker on the established position. Lay the cover with
the hole on the sample beaker.
3 Fasten a membrane filter to the Luer adapter of the transfer robotic
arm.
4 Move the lift to the Target position = Work position. The filter
must rest on the cover of the sample beaker. Correct the lift position
if necessary.
5 If the position has had to be corrected, switch to the Assign posi-
tion tab.
6 Select the Work position for = Special beaker 1 in the Lift posi-
tion field.
7 Click on the associated [Assign] button.
8 Switch back to the Move tab.
9 In the manual control in the left-hand window, click on the Tower 2
item under 815_1 (815 Robotic …) and then select the Move tab.
10 Under Rack position, enter the Target position = Special
beaker 2 and click on [Start].
11 Place a suitable sample beaker on the established position and fill it
with a sponge or a crushed absorbent paper towel. This is to be used
to dab off the moist Polytron aggregate.