Figure 31, Install pump tubing connection with filter – Metrohm 850 Professional IC Anion MCS Prep 3 User Manual

Page 58

background image

2.15 Peristaltic pump

■■■■■■■■■■■■■■■■■■■■■■

48

■■■■■■■■

850 Professional IC –Anion – MCS – Prep 3

3 Connecting the pressure side

Note

Depending on the use of the peristaltic pump, on the pressure side
you can either connect:

Case A: a 6.2744.180 pump tubing connection with filter
(see Figure 31, page 48) or

Case B: a 6.2744.160 pump tubing connection without filter
(see Figure 32, page 49).

For pumping the auxiliary solutions to the MSM or to the SPM, a
6.2744.180 pump tubing connection with filter must be used.

Case A: 6.2744.180 pump tubing connection with filter:

1

2

3

Figure 31

Install pump tubing connection with filter

1

Union nut

2

Adapter

3

Tubing olive with filter holder

Slide union nut (31-1) onto the pump tubing.

Select a suitable adapter (31-2) and slide it onto the pump tubing.
The type of adapter depends on the pump tubing (see Table 1,
page 49)
.

Place the tubing olive with filter holder (31-3) onto the pump tub-
ing.

Screw the union nut (31-1) onto the tubing olive (31-3).

or

Case B: 6.2744.160 pump tubing connection without filter: