Figure 31, Install pump tubing connection with filter – Metrohm 850 Professional IC Anion MCS Prep 3 User Manual
Page 58

2.15 Peristaltic pump
■■■■■■■■■■■■■■■■■■■■■■
48
■■■■■■■■
850 Professional IC –Anion – MCS – Prep 3
3 Connecting the pressure side
Note
Depending on the use of the peristaltic pump, on the pressure side
you can either connect:
■
Case A: a 6.2744.180 pump tubing connection with filter
(see Figure 31, page 48) or
■
Case B: a 6.2744.160 pump tubing connection without filter
(see Figure 32, page 49).
For pumping the auxiliary solutions to the MSM or to the SPM, a
6.2744.180 pump tubing connection with filter must be used.
Case A: 6.2744.180 pump tubing connection with filter:
1
2
3
Figure 31
Install pump tubing connection with filter
1
Union nut
2
Adapter
3
Tubing olive with filter holder
■
Slide union nut (31-1) onto the pump tubing.
■
Select a suitable adapter (31-2) and slide it onto the pump tubing.
The type of adapter depends on the pump tubing (see Table 1,
page 49).
■
Place the tubing olive with filter holder (31-3) onto the pump tub-
ing.
■
Screw the union nut (31-1) onto the tubing olive (31-3).
or
Case B: 6.2744.160 pump tubing connection without filter: