2.14 Sample preparation module (SPM)

■■■■■■■■■■■■■■■■■■■■■■

44

■■■■■■■■

850 Professional IC –Anion – MCS – Prep 3

Weak liquors can directly be forwarded to the SPM as with cation
exchange. However, strong liquors (e.g. 30 % NaOH) could exceed the
capacity of the SPM. In order to ensure a sufficient capacity for exchang-
ing the Na

+

ions for H

+

ions in all cases, only a small part of the sample

(e.g. 20 µL) is therefore given onto the SPM. For measuring the sample
the sample loop of an additional injection valve(26-A) (e.g. on the
Sample Processor) is needed.

Procedure of the neutralization with the SPM

1 Measuring the sample

Valve positions:

Injection valve 1 (26-A) to position Fill.
– Injection valve 2 (in the instrument) in position Inject.

On the injection valve 1 (26-A) the sample is carried through the
sample loop and is measured.

1

A

B

Figure 26

Measuring the sample

1

Sample

A

Injection valve 1
With sample loop. For measuring the sam-
ple.

B

Injection valve 2
With preconcentration column. For precon-
centrating and injecting the analyte ions.

2 Neutralization and preconcentration

Switch both injection valves:

Injection valve 1 (27-A) in position Inject.
Injection valve 2 (27-C) to position Fill.

The sample is carried with the transfer solution (27-2) onto the
SPM (27-B).
Na

+

ions are exchanged for H

+

ions.

The analyte ions and the transfer solution flow to the injection
valve 2.