beautypg.com

12 injection valve, 1 protection, Protection – Metrohm 850 Professional IC Anion MCS Prep 3 User Manual

Page 109: Figure 53, Pump tubing connection – changing the filter

background image

■■■■■■■■■■■■■■■■■■■■■■

4 Operation and maintenance

850 Professional IC –Anion – MCS – Prep 3

■■■■■■■■

99

1

2

3

Figure 53

Pump tubing connection – Changing the filter

1

Tubing olive

2

Filter 6.2821.130
Packaging contains 10 items.

3

Filter housing

Replacing the filter

1 Unscrewing filter screw

Screw the filter screw (53-3) out of the tubing olive (53-1) with
the aid of two 6.2621.000 adjustable wrenches.

2 Replacing the filter

Remove the old filter (53-2) with tweezers.

Place the new filter (53-2) flat in the tubing olive (53-1) with
tweezers.

3 Mounting filter screw

Screw the filter screw (53-3) back into the tubing olive (53-1) and
tighten by hand. Then additionally tighten with two 6.2621.000
adjustable wrenches.

4.12

Injection valve

4.12.1

Protection

To prevent contamination of the injection valve, a 6.2821.120 inline filter
(see Chapter 2.11, page 36) should be mounted between the high pres-
sure pump and the pulsation damper.